Human IL-6 ELISA Kit
Validation Data Gallery
Product Information
KE00385 is a solid phase sandwich Enzyme Linked-Immuno-Sorbent Assay (Sandwich ELISA). The human IL-6 ELISA kit is to be used to detect and quantify protein levels of endogenous human IL-6. The assay recognizes human IL-6. An antibody specific for human IL-6 has been pre-coated onto the microwells. The human IL-6 protein in samples is captured by the coated antibody after incubation. Following extensive washing, another horseradish peroxidase (HRP)-conjugated antibody specific for human IL-6 is added to detect the captured human IL-6 protein. For signal development, followed by Tetramethyl-benzidine (TMB) reagent. Solution containing sulfuric acid is used to stop color development and the color intensity which is proportional to the quantity of bound protein is measurable at 450 nm with the correction wavelength set at 630 nm.
| Product name | Human IL-6 ELISA Kit |
| Tests | 1 X 96 well plate |
| Sample type | Serum, Plasma, Cell culture supernatant 备注:对未验证样本,我们提供技术支持和售后保障。 |
| Assay type | Sandwich |
| Sensitivity | 0.2 pg/mL |
| Range | 7.8-500 pg/mL |
| Reactivity | Human |
| Tested applications | Sandwich ELISA |
| Gene ID (NCBI) | 3569 |
Recovery
| Sample Type | Average | Range |
|---|---|---|
| Cell culture supernatant | 90% | 86%-99% |
| Human serum | 82% | 72%-98% |
IntraAssay
| Sample | n | mean ( pg/mL) | SD | CV% |
|---|---|---|---|---|
| 1 | 20 | 229.9 | 9.0 | 3.9 |
| 2 | 20 | 52.2 | 1.7 | 3.3 |
| 3 | 20 | 15.2 | 0.8 | 5.2 |
InterAssay
| Sample | n | mean ( pg/mL) | SD | CV% |
|---|---|---|---|---|
| 1 | 24 | 242.1 | 11.3 | 4.6 |
| 2 | 24 | 55.5 | 1.8 | 3.2 |
| 3 | 24 | 15.7 | 0.9 | 5.5 |
Background Information
Interleukin-6 (IL6) is an interleukin that acts as both a pro-inflammatory and anti-inflammatory cytokine. IL6 protein is secreted by a variety of cell types including T cells and macrophages as phosphorylated and variably glycosylated molecule. IL6 Plays an essential role in the final differentiation of B-cells into Ig-secreting cells involved in lymphocyte and monocyte differentiation. It induces myeloma and plasmacytoma growth and induces nerve cells differentiation Acts on B-cells, T-cells, hepatocytes, hematopoietic progenitor cells and cells of the CNS. IL6 is also considered a myokine, a cytokine produced from muscle, and is elevated in response to muscle contraction. IL6 has been shown to interact with interleukin-6 receptor and glycoprotein 130. Additionally, IL6 is involved in hematopoiesis, bone metabolism, and cancer progression, and has been defined an essential role in directing transition from innate to acquired immunity.
Properties
| Storage Instructions | All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. |
| Synonyms | IL6,IL-6,B-cell stimulatory factor 2,IFN-beta-2,IL |
FAQ
1、我的样本稀释液有限,如何处理稀释倍数较大的样本?
我们建议您对样本进行梯度稀释。例如,人血清样本需要1000000倍稀释时,您可以分3个梯度,每个梯度100倍,进行稀释:
第一步:取2 uL人血清样本,加入198 uL样本稀释液,混匀得到100倍稀释的人血清样本
第二步:取2uL 100倍稀释的人血清样本+ 198 uL样本稀释液进行稀释,混匀得到10000倍稀释的人血清样本
第三步:取2uL 10000倍稀释的人血清样本+ 198 uL样本稀释液进行稀释,混匀得到1000000倍稀释的人血清样本。
如果样本需要进一步稀释,您可以使用更多梯度。除了我们推荐的样本稀释液,不建议您使用其他溶液或者水稀释。如果您的样本稀释液仍然不足,请联系我们的客服为您提供更多样本稀释液。
Publications
| Species | Sample Type | Title |
|---|---|---|
Am J Transl Res Comprehensive evaluation of combined AngioJet and drug-coated balloon therapy for diabetic foot ulcers with lower extremity arterial occlusive disease | ||
Int Endod J Laboratory Investigation of Lipid Metabolic Reprogramming in Pulpitis: The Linolenic Acid-FASN-NR4A1 Axis in Modulating Dental Pulp Inflammation. | ||
J Endocrinol Invest Puerarin suppress adipogenesis and inflammation on orbital fibroblasts from patients with thyroid eye disease. | ||
Sci Rep Sepsis-associated skeletal muscle wasting is ameliorated by pharmacological inhibition of the STAT3 signaling pathway in mice. | ||
Cell Death Dis K48 and K63 linkage-competed ubiquitination of BECN1 promotes circPDE4D-mediated autophagy in chronic obstructive pulmonary disease. | ||
Ther Apher Dial Efficacy and Safety Analysis of Roxarestat in Regulating Renal Anemia in Patients on Maintenance Hemodialysis. |