Speedy™ Human BCL11A One-Step ELISA Kit
Sensitivity
0.02 ng/mL
Range
0.156-10 ng/mL
Reactivity
Human
Cat no : SE50110
Synonyms
B cell lymphoma/leukaemia 11A, B-cell CLL/lymphoma 11A, B-cell lymphoma/leukemia 11A, BCL-11A, COUP-TF-interacting protein 1
Validation Data Gallery
Product Information
| Product name | Speedy™ Human BCL11A One-Step ELISA Kit |
| Tests | 1 X 96 well plate |
| Sample type | Serum, Plasma |
| Assay type | Sandwich |
| Sensitivity | 0.02 ng/mL |
| Range | 0.156-10 ng/mL |
| Reactivity | Human |
| Tested applications | Sandwich ELISA |
| Gene ID (NCBI) | 53335 |
Recovery
| Sample Type | Average | Range |
|---|---|---|
| Human plasma | 89% | 75%-114% |
IntraAssay
| Sample | n | mean ( ng/mL) | SD | CV% |
|---|---|---|---|---|
| 1 | 8 | 5.82 | 0.51 | 8.76 |
| 2 | 8 | 1.31 | 0.04 | 3.05 |
| 3 | 8 | 0.71 | 0.04 | 5.63 |
InterAssay
| Sample | n | mean ( ng/mL) | SD | CV% |
|---|---|---|---|---|
| 1 | 16 | 5.89 | 0.52 | 8.83 |
| 2 | 16 | 1.36 | 0.09 | 6.62 |
| 3 | 16 | 0.71 | 0.03 | 4.23 |
Background Information
B-Cell Lymphoma/Leukemia 11A(BCL11A) is a highly conserved zinc finger gene. It acts as a myeloid and B-cell proto-oncogene, and may play key roles in leukemogenesis and hematopoiesis. It is an essential factor in lymphopoiesis, and required for B-cell formation in fetal liver. May function as a modulator of the transcriptional repression activity of ARP1. It is also involved in lymphoid malignancies through translocation or amplification events. It has recently been shown that BCL11A expression is inversely correlated with γ globin expression.
Properties
| Storage Instructions | All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. |
| Synonyms | B cell lymphoma/leukaemia 11A, B-cell CLL/lymphoma 11A, B-cell lymphoma/leukemia 11A, BCL-11A, COUP-TF-interacting protein 1 |
FAQ
1、我没有恒温振荡器,这会对实验时间和结果产生影响吗?
如果您没有恒温振荡器,完成这个实验大约需要 2.5 小时。在加入抗体混合物后,我们建议您将板条37℃下静置孵育2小时。在加入 TMB 后,您只需让板条37℃下静置孵育15分钟即可。我们的研发人员通过对比实验已经证明,尽管信号会略有降低,但您仍然可以获得有效的实验数据。